Brand Name : BonninTech
Model Number : AutoPure-32, AutoPure-96
Certification : ISO
Place of Origin : China
MOQ : 1 set
Price : USD8000-25800/set
Payment Terms : L/C, T/T, Western Union, MoneyGram
Supply Ability : 500 set/year
Delivery Time : 5-8 working days
Packaging Details : Standard export wooden box
Equipment : Nucleic Acid Extraction System
Sample capacity(wells) : 32/96
Process volume : 50~1000ul
AutoPure-32 AutoPure-96 Nucleic Acid Extraction and Purification System
Auto-Pure32 is a bench top device for DNA and RNA purification by using magnetic separation technology from different
starting materials, such as blood, cultured cells or bacteria, tissues, cell-free body fluids and plant samples. The device is capable of using
magnetic rods to transfer particles through the various purification phases of binding, mixing, washing and elution, offering a solution with minimized hands-on time. The DNA and RNA purification is of high quality and high molecular weight.
Features
1. It is easy to use with 7 inch touch screen
2. Very simple operation (easy to install, operate, maintain) without computer
3. Very fast extraction protocol, 15~40 minutes / cycle depending on sample type
and method
4. Universal built-in program for easy using
5. High purity and excellent yield of nucleic acid
6. UV lamp to avoid cross contamination
7. 3 shortcut key to make for easy running, stopping the magnetic beads program
8. Open system can optimize purification proposal according to various magnetic
beads kits
9. Drawer design to prevent possible injuries
10. With special plastic consumables to avoid cross contamination
11. Improves workflow, and allows staff to perform other value-added tasks
12. Ensures impurities are removed; improved sample quality leads to better
downstream analyses
13. Capable of extracting 1~20 samples or 1~32samples per run
14. Alarm for indicating the completion of purification
15. Pause function for emergent stop
Technical Data
Type | Auto-Pure32 |
Throughput | 1~32 |
Sample volume | 20~1000ul |
Collection efficiency of the magnetic particles | >95% |
Magnetic rod number | 32 |
Plate types | 96 Deepwell plate |
Extraction steps | Lysis, Sample binding, Washing and elution |
Purification accuracy | 100 copy sample positive rate >95% |
Heating Temperature | For cell lysis: Ambient temperature ~90℃ |
For nucleic acid elution | Ambient temperature ~90℃ |
Operation | 7 inch color touch screen |
Storage Capacity | more than 100 programs |
Protocol Management | Create, edit , delete, protocol mode |
Pollution Control | UV light |
Extension interface | 4 standard USB port, built-in SD card |
Exhaust | Fan |
Power Supply | AC110/220V, 50/60Hz |
Dimensions | Dimensions |
Weight | 23kg |
Accessories
Code | Description |
AS-17021-01 | Megnetic rod’s tip for Auto-Pure32 |
AS-17021-02 | 96-Deepwell plate for Auto-Pure32 |
Auto-pure96 Nucleic Acid Extraction System
Type | AutoPure-96 |
Throughput | 1~96 |
Process volume | 50~1000ul |
Collection efficiency | >95% |
Magnetic rod number | 96 |
Purification accuracy | 100 copy sample positive rate>95% |
Stability | CV<5% |
Plate types | 96 Deepwell plate |
Healting for lysis tube | Ambient temperature~120°C |
Heating for elution tube | Ambient temperature~120°C |
Operation | 7 inch color touch screen |
Extraction steps | Lysis, Sample binding, Washing and elution |
Storage Capacity | more than 100 programs |
Protocol Management | Create, edit, delete, protocol mode |
Pollution Control | UV light |
Lighting | Yes |
Extension interface | 4 standard USB port, built-in SD card |
Exhaust | Fan |
Power Supply | 450W |
Dimensions | 560×620×500mm |
Weight | 54Kg |
Accessories
Code | Description |
AS-17060-00 | Auto-Pure96 Nucleic Acid Purification System, AC120V/240V, 50/60Hz |
AS-17061-01 | Magnetic tip comb for Auto-Pure96 |
AS-17061-02 | 96-Deepwell plate for Auto-Pure96 |
AS-17061-03 | 96-Elution plate for Auto-Pure96 |
Operational advantages
1. High-throughput:
Auto-Pure96 can process 96x1ml samples simultaneously
2. Open and multiple functions design can be provided more methods for reagent optimization, and also a reliable choice
3. Touch screen interface and easy operation, parameters can be tuned based on experimental requirements
4. Accurate temperature control system ensures stability of purification performance
5. Patent structure design with center rotary table which makes it occupies smaller area of magnetic rod tip comb, which guaranteed stability and liability (Success rate >99.99%)
Gentier96E Real Time PCR System for Covid-19 Testing
Gentier 96E | Gentier 96S | Gentier 96R | Gentier 96C | |
Thermal Block | ||||
Sample capacity(wells) | 96 | |||
Reaction Volume | 0-100 μL | |||
Consumables | 0.2ml 96-well Plates(unskirt); 0.2ml 8-strip tubes ,0.2ml PCR single tube(Optical flat cap,clear,frosted,white tube) | |||
Temperature range | 0°C-100°C | |||
Heating/cooling method | Peltier | |||
Max heating rate | 6.1°C/sec | |||
Average heating rate | 4.5°C/sec | |||
Max cooling rate | 5.0°C/sec | |||
Average cooling rate | 2.8°C/sec | |||
Temperature accuracy | ±0.1℃ | |||
Temperature uniformity | ±0.1℃ | |||
Gradient Range | 1°C-40°C | N/A | 1°C-40°C | N/A |
Gradient block | 12 row | 12 row | ||
Special temperature portocol | Gradient PCR,Long PCR,Touch Down PCR | Long PCR,Touch Down PCR | Gradient PCR,Long PCR,Touch Down PCR | Long PCR,Touch Down PCR |
Heat Lid | ||||
Temperature range | Room Temperature–110°C | |||
Optical System | ||||
Excitation Source | 6 LEDs | 4LEDs | ||
Detector | Photodiode | |||
Detection Position | Excitation and scan at top | |||
Detection method | 6 channels scanning at the same time, no edge effect. | |||
Detection time | 7 seconds for 96 wells for all channels | |||
Range of excitation/ emission wavelengths (nm) | 1 465 /510 (FAM,SYBR Green I,SYTO9,EvaGreen,LC Green) 2 527/ 563 (HEX,VIC,TET,JOE) 3 580 /616 (ROX,Texas Red) 4 632/ 664( Cy5) 5 680 /730 (Alexa Fluor680) 6 465 /616 (FRET) | 1 465 /510 (FAM,SYBR Green I,SYTO9,EvaGreen,LC Green) 2 527/ 563 (HEX,VIC,TET,JOE) 3 580 /616 (ROX,Texas Red) 4 632/ 664( Cy5) | ||
Probe | Taqman Probe,Molecular beacons probe,Scorpion probe ,FRET | Taqman Probe,Molecular beacons probe,Scorpion probe | ||
Multiplexing | Up to 6 targets | Up to 4 targets | ||
Fluorescence Linearity | r≥0.990 | |||
Fluorescence Dynamic Range | Adjustable | |||
Performance | ||||
Sample Linearity | /r/≥0.999 | |||
Sample repeatability | Ct value CV≤0.5% | |||
Sample Dynamic Range | 1-1010 copies | |||
Software Funtions | ||||
Data Analysis modes | Qualitative analysis,Absolute quantification, Relative
quantification,Genotyping analysis,Endpoint analysis, Melt curve analysis, High Resolution Melting | |||
Control modes | 1.10.4' touch screen 2. PC direct control 3. WLAN control(One PC can control max 10 units, and device can be controled by any PC in the WLAN) | |||
Sample Drawer | Touch sreen commend | |||
Data Stroage | Upload and download through USB disk,1000 results can be stored in machine | |||
Power failure protection | Automaticly start running experiments after power supply, no need wait PC software | |||
Customize Report | Templetes reserved ,report can customized | |||
Administration Management | Administrator can set funtions limts for users | |||
Transport Locker | Automaticly detect transport locker | |||
Fault Management | Fault report and analysis, solution instruction | |||
LIS connection | CSV,Excel,TXT format data output,open port for LIS connection | |||
Others | ||||
PC operating system | Win 7, Win 10 | |||
Communication Port | 1 ethernet and 3 USB | |||
Footprint(WxDxH) | 355mm X 480mm X 485mm | |||
Weight | 30kg | |||
Power usage | AC 100 to 125 V / 200 to 240 V (50/60 Hz). | |||
power consumption | 900VA | |||
Work environment | Tempereture:10°C~30°C Humidity:20%~85% |
![]() |